Schematic Board & Software TriStar II 3020Service
2-8 Jul 09
Schematic Board & Software
The schematic diagram of the 3020 is displayed on the computer screen by the TriStar
software. The schematic shows:
The current state of each valve.
The pressure reading from the transducers.
The temperature reading in the manifold.
Location and status of the elevator.
The software is designed to work on computers running Windows XP or Windows Vista. The
software also supports importing data from the TriStar 3000 and Gemini for analysis and
comparison with TriStar II 3020 plots. For more instructions on how to use the software and
run the TriStar analyzer, refer to Chapter 3 - Operational Procedures in the Operators
Manual.
1
2
3
4
5
6
7
8
9
10
11
12
1 - Gas Inlet Manifold
2 - Vacuum Pump
3 - Servo Valve
4 - Manifold Transducer
5 - Manifold Thermocouple
6 - Port 1 Transducer
7 - Port 3 Transducer
8 - Port 2 Transducer
9 - P
0
Transducer
10 - Elevator
11 - Status Bar
12 - 10-Torr Transducer (Krypton only)
All transducers are 1000-Torr unless otherwise stated.