Troubleshooting VacPrep 061
3-2 Feb 2013
Table 3-1. Common Operational Problems (continued)
What Happened Why What To Do
Gas seems to be leaking
from gas bottle.
The flow control valves were
left in the open (GAS) position
when stations were not in use.
Keep the flow control valve
closed (OFF) at on all stations
that are not being used for
degassing.
The gas flow rate is set too
high.
Decrease the flow rate by
reducing the regulator
pressure.
A flow control valve is
leaking.
Call the appropriate service
personnel.
No gas flow when the flow
control valve is opened.
Sample material in the gas
delivery tube is blocking gas
flow.
Clean the gas delivery tube;
replace if necessary. Refer to
Cleaning and Replacing Gas
Delivery Tubes, page 3-6.
The flexible tubing is twisted. Remove twist, replace the
flexible tubing if necessary.
Refer to Replacing Flexible
Tubing, page 3-7.
Gas pressure is set too low. Increase the flow rate by
raising the regulator pressure.
Gas bottle is empty. Replace the gas bottle.
Valve inlet tubing is detached. Check the valve inlet tubing
connections. Refer to
Connecting or Replacing
Valve Inlet Tubing, page 3-9.
Flow control valve is
damaged.
Call the appropriate service
personnel.