© Midmark Corporation 2020
English - 10
TP202 20-42-FO-00014 Rev A1 C2169
003-1879-99
VA1233
Handpieces
The delivery system controls up to four handpieces. The handpiece pressure is controlled by handpiece
drive air adjustment valves. The gauge on front of the delivery system indicates the handpiece pressure.
Handpiece pressure should not exceed 60 PSI under normal operation. The scaler tool (optional) has an
adjustment knob for the intensity level.
Operation
Delivery System
Air/Water Syringe
There are air and water buttons on top of the syringe. When both buttons are depressed you will get a
mist of both water and air. Air and water adjustments are located under the syringe holder. Insert flat head
screwdriver into adjustment screw(s). Counter-Clockwise to increase flow and Clockwise to decrease flow.
Turn screw until desired air/water flow is coming through the instrument.
Gauge
Air
Water
Equipment Alert
Refer to handpiece
manufacturer for suggested PSI
settings. Only check one handpiece
at a time.
Handpiece
Holders
Scaler
Adjustment
Air
Water
Caution
Handpieces and Syringe Tip should be changed
or disinfected between each patient.
Drill Polisher
Syringe
Scaler