Operation
Water System
Water Bottles
The unit has two 1 liter bottles. The water bottles supply water to the handpieces, air/water syringe and a
quick disconnect on the back of the cart. Air flows through the water regulator and the toggle switch to the
water bottles. The water system is regulated to 40 PSI. Use the water needle valve to adjust water flow
for each instrument. The toggle switch between the water bottles that controls which bottle will supply the
water. When the toggle is in the center neither water bottle is pressurized. Move the toggle toward the
left to pressurize the left bottle and depressurize the right bottle. Toggle to the right to pressurize the right
bottle and depressurize the left bottle.
Toggle
To fill water bottles...
A) Move toggle to center.
B) Remove water bottle and fill. (Distilled Water Is Recommended)
C) Replace water bottle and move toggle toward filled water bottle.
Note: Water bottles should be checked in between each patient and filled if necessary.
Water Quick Disconnect
The water quick disconnect has a blue ring. You can attach accessories to the cart that will utilize the
water from the cart. You may need to add a regulator if attaching equipment that can not handle 40 PSI.
Water
Needle
Valve
Water
Regulator
Water
Bottles
Water Quick
Disconnect
Caution
Failure to properly seal water bottles can lead to patient injury.