© Midmark Corporation 2020
English - 23
TP202 20-42-FO-00014 Rev A1 C2169
003-1879-99
Midmark High-Speed Handpiece
Cleaning/Maintenance
Equipment Alert
All Midmark Handpieces must be sterilized and lubricated after every patient. Lubrication will elongate life,
increase performance and eliminate debris from within handpiece. Lubrication can be performed between
handpiece cleanings if the handpiece appears to be under performing. If equipped, the swivel coupler
can be sterilized in a bag at the same temperatures as a handpiece.
1) Do not use Flash sterilization techniques or damage to components will occur, voiding warranty.
2) Never submerge components in any cleaning or disinfecting solution.
3) Do not use Ultrasonic Cleaners on handpieces or couplers.
4) Do not Autoclave or Chemiclave for extended / Unusual periods of time (such as overnight).
5) Do not leave handpiece components in sterilizer after cycle is completed.
WARNING
Handpiece should be sterilized between patients to prevent cross-contamination
which could result in serious illness or death from infectious diseases.
To clean the water port(s)...
A) Use water spray tool to periodically clean spray ports as needed.