© Midmark Corporation 2020
English - 24
TP202 20-42-FO-00014 Rev A1 C2169
003-1879-99
Handpiece Connections
To clean the handpiece...
A) Detach handpiece from hose coupler (if equipped with a swivel quick disconnect
remove handpiece then remove swivel from coupler with wrench).
B) Remove bur from handpiece.
C) Keeping the head of handpiece upright, clean external surface of handpiece
thoroughly to remove saliva, blood, and other organic soil. Scrub handpiece with
a soft brush and warm water to remove debris. Rinse and dry.
Bur Removal
Hose Coupler
Fiber Optic
Module
Swivel Coupler
(Quick Disconnect)
Swivel Coupler
Wrench
Equipment Alert
Line up components properly to prevent
damage resulting in hand piece failure.
Midmark High-Speed Handpiece
Cleaning/Maintenance continued...