TP202 20-42-FO-00014 Rev A1 C2169
003-1879-99
Drive Air Port -
2-Hole 5-Hole4-Hole
To clean/condition the handpiece drive air ports...
(Applicable on Non-Swivel High-Speed Handpieces Only).
A) Assemble spray tube onto spray can head by firmly pushing tube into conditioner
can spray tip as far as possible.
B) Insert tube into drive air port of handpiece.
C) Direct handpiece away from your body and towards an absorbent towel. Depress
handpiece conditioner can top to spray until conditioner flows from both ends of
handpiece head (1 to 2 seconds).
D) If overspray on towel appears discolored repeat spray for an additional two seconds.
Remove handpiece and wipe excess conditioner from handpiece with a clean dry
towel.
Spray Tube
Midmark High-Speed Handpiece
Cleaning/Maintenance continued...
Apply Cleaner/
Conditioner Here