At the beginning of each work day...
• Fill water bottles with fresh distilled water.
• Flush handpieces.
Between each patient...
• Remove disposable tips,
instruments, etc..
• Flush delivery system.
• Disinfect handpieces and scaler
per manufacturers instructions.
• Disinfect syringe tip.
• Fill water bottles.
To Flush Delivery System...
A) Remove all handpieces that use water.
B) Open water needle valve.
C) Hold end of tubing over a container and depress foot control.
D) Allow water to run 20 to 30 seconds.
E) Install newly disinfected handpieces and syringe tip.
F) Refill water bottle.
Water Needle Valve
Note
Distilled water is not mandatory.
However, water must meet
requirements for drinking.
Equipment Alert
AUTOCLAVABLE SYRINGE TIP STERILIZATION
The autoclavable syringe tips supplied with the delivery system
must be sterilized prior to use with each patient, including initial use. Be sure to
thoroughly rinse and clean syringe tips prior to sterilization, any debris may reduce
the effectiveness of the sterilization. Recommended sterilization process is steam
autoclave. Recommended parameters are 125°C (250°F) and 106 kPa (15 PSI) for 40
minutes at temperature and pressure.
Equipment Alert
HANDPIECE STERILIZATION
Between each procedure complete
manufacture’s instructions for “Infection Control /
Sterilization found in the documentation included
with the handpieces. Failure to do this will cause
premature failure of handpieces.
Midmark 1000 Cleaning/Maintenance continued...