5.1. Safety precautions
The personnel involved in the decontamination procedure of PrestoCHILL and its optional (if present) must
wear the appropriate individual protective devices, in particular:
• Single use, HI-RISK protective gloves (100% hi-risk nitrile or equivalent). Individual protection device
that meets the standard EN 374 and the Directive 89/686/CEE.
• Antibacterial mask face for surgery room and white coat. Individual protection device, meets the
standard EN 149 Grade FFP3D and the Directive 89/686/CEE.
5.2. PrestoCHILL decontamination and cleaning procedure
5.2.1. Detergents
The following instructions must be strictly complied with in order to perform proper decontamination and
cleaning procedure.
For further information, contact our specialists at the following email address:
application@milestonemedsrl.com.
Before any operation, disconnect from the power supply mains. Do not use direct or high
pressure water to clean the instrument.