User Manual
Elix® 20, 35, 70 and 100

Table of Contents

Questions and Answers

  • M
    michaelphillipsSep 23, 2025
    What to do if Millipore Water Filtration Systems says bad Progard type?
    • C
      Carrie BerrySep 23, 2025
      If your Millipore Water Filtration System displays 'BAD PROGARD TYPE', check the type of pack being installed. For further assistance, contact Millipore SAS.
  • K
    Kristina CannonAug 23, 2025
    How to clear ASM UV lamp exchange overdue message on Millipore Elix 20 Water Filtration Systems?
    • P
      Paul WaltonAug 25, 2025
      The 'ASM UV LAMP EXCH. OVERDUE: XX DAYS' message on your Millipore Water Filtration System will disappear once a new ASM UV Lamp is installed and its software timer is reset.
  • K
    Kimberly RobinsonAug 25, 2025
    How to clear EXCH. UV lamp in XX days message on Millipore Elix 20?
    • O
      oconnorkristaAug 26, 2025
      The message 'EXCH. UV LAMP IN XX DAYS' on your Millipore Water Filtration System will disappear when you install a new UV Lamp and reset the software timer.
  • D
    Danielle PenaAug 27, 2025
    What to do if distilled water conductivity is less than mini on Millipore Water Filtration Systems?
    • R
      Robert DeleonAug 27, 2025
      If the Millipore Water Filtration System displays 'DIST WATER CONDUCTIVITY < MINI', keep the loop operating for a few minutes to remove any air that could be trapped in the Resistivity Sensor. If problems persist, contact Millipore SAS for assistance.
  • B
    Bradley OchoaAug 27, 2025
    What to do if the Millipore Water Filtration Systems shows a low feed pressure?
    • M
      micheal10Aug 27, 2025
      If the Millipore Water Filtration System displays 'LOW FEED PRESSURE', first ensure the feedwater supply is turned on. Next, check if the Progard TL Pack is clogged by observing the RO reject water in FLUSH mode. If the water pressure issue is resolved during FLUSH mode, the alarm will automatically reset.
  • C
    chadmoralesSep 17, 2025
    What to do if Progard TL pack type error appears on Millipore Elix 20 Water Filtration Systems?
    • G
      guerrachristopherSep 17, 2025
      If your Millipore Water Filtration System displays 'PROGARD TL PACK TYPE ERROR', check the catalogue number of the Progard TL using its label and replace it if necessary.
  • B
    Benjamin MullenSep 14, 2025
    How to fix TOC meter error 2 on Millipore Elix 20?
    • R
      Ronald JohnsSep 14, 2025
      If you encounter a 'TOC METER ERROR 2' on your Millipore Elix System, try turning the system off and then on again. Then, dispense water for several minutes. If the error persists, contact Millipore SAS for further assistance.
  • P
    paigewilliamsSep 15, 2025
    How to fix TOC meter error 3 on Millipore Elix 20 Water Filtration Systems?
    • M
      maxwellcarlsonSep 16, 2025
      If you encounter a 'TOC METER ERROR 3' on your Millipore Elix System, try turning the system off and then on again. Then, dispense water for several minutes. If the error persists, contact Millipore SAS for further assistance.
  • M
    Maurice MurphySep 6, 2025
    What to do for high pump pressure on Millipore Elix 20 Water Filtration Systems?
    • F
      floresedwardSep 6, 2025
      To address a 'HIGH PUMP PRESSURE' issue on your Millipore Water Filtration System, try setting the RO Pump Pressure lower. If the problem persists, contact Millipore SAS for additional assistance.
  • L
    Laura WoodSep 3, 2025
    How to fix Progard TL pack not in place error on Millipore Water Filtration Systems?
    • N
      nmaciasSep 3, 2025
      To resolve the 'PROGARD TL PACK NOT IN PLACE' error on your Millipore Water Filtration System, make sure the Progard TL is installed with its “TAG” at the bottom.

Summary

Introduction

Safety Information

Important safety statements, warnings, and symbols for system operation.

Product Information

Using the Elix System

Maintenance

Scheduled Maintenance Table

Summary of typical maintenance tasks, frequency, and procedures.

How to Replace the Progard TL

Step-by-step guide for replacing the Progard TL pretreatment pack.

How to Replace the UV Lamp Inside the Elix System

Instructions for replacing the UV lamp within the Elix system.

How to Sanitise the RO Cartridge(s)

Methods for sanitizing the RO cartridges, both automatic and manual.

How to Clean the RO Cartridge(s)

Procedures for cleaning RO cartridges using specific cleaners.

How to Replace the QGard TL1 Polisher Pack

Step-by-step guide for replacing the QGard TL1 Polisher Pack.

How to Clean the A10 TOC Monitor

Procedures for cleaning the A10 TOC Monitoring Device.

Troubleshooting

Maintenance, Alarm, Alarm Stop and Custom Messages

Classification and explanation of different system messages and indicators.

An Alarm Message or Maintenance Message is Displayed

Detailed list of LCD messages, their meanings, and recommended actions.

Millipore Elix 20 Specifications

General IconGeneral
TypeWater Purification System
ModelElix 20
Output Flow Rate20 L/h
Power Supply100-240 V, 50/60 Hz
Feed Water RequirementsPotable water