8
Accutorr
®
Quick Reference Guide
Code Description Solution
26 Temp module error
Restart the monitor. If the error persists, contact
Mindray's Technical Service department.
27 Temp probe error
Verify that the probe is in the probe well, or cool
the probe and re-insert into the probe well.
28
Ambient temp
overrange
Change or cool the environment and retry.
29 Temp overrange Contact Mindray's Technical Service department.
40
Power board comm.
error
Restart the monitor. If the problem persists,
contact Mindrays Technical Support department.