7
Accutorr
®
Quick Reference Guide
ERROR CODE
Code Description Solution
01 NIBP overrange Contact Mindray's Technical Service department.
02 NIBP module error
Restart the monitor. If the error persists, contact
Mindray's Technical Service department.
03 NIBP comm. error
04 NIBP air pressure error
Check for loose cu or pinched hose. Restart the
monitor and retry. If the error persists, contact
Mindray's Technical Service department.
05 NIBP weak signal
Check the patient’s condition and change the
cu application site. If the error persists, replace
the cu.
06 NIBP excessive motion
Check the patient’s condition and reduce patient
motion.
07 NIBP cu overpressure
Check for kinked hose or excessive patient
movement. Restart the monitor and try again.
08 NIBP illegally reset Contact Mindray's Technical Service department.
09 NIBP timeout
Check the patient’s condition and NIBP
connections, or replace the cu.
10 NIBP cu type wrong
Verify the patient size and use the appropriate
cu.
11 NIBP air leak
Ensure the cu is applied correctly and check the
ttings for leaks. If the problem persists, contact
Mindray's Technical Service department.
17
SpO2 board fault
(Masimo)
Discontinue monitoring SpO2 and contact
Mindray's Technical Service department.
18 SpO2 module error
Restart the monitor. If the error persists, contact
Mindray's Technical Service department.
19 PR overrange Contact Mindray's Technical Service department.
20 SpO2 low perfusion Move the sensor to a site with better perfusion.