4-3
System
Fault
Code
System
Exception
Code
Alarm Description Possible Cause
Handling
Method
SE101 /
AC-DC Power Output
Too High
1.The AC power
supply is
abnormal.
2.The AC power
cord is poorly
connected or
damaged.
3.The AC-DC
module is faulty
or abnormal.
4.The control
board is faulty.
1.Check if the
AC power supply
meets the
specifications.
2.Reconnect or
replace the AC
power cord.
3.Try to restart
the equipment. If
it is not
recovered,
replace the
AC-DC module.
4.Replace the
control board.
/ SA102
AC-DC Power Output
Too Low
SE103
/
EDC Too High
1.The control
board is faulty.
2.The system
may be
short-circuited.
3.The specified
serial cable is not
used, causing
unintended
power to enter
the system, alarm
SA104. (In this
case, the alarm
indicates no
actual fault. It is
unnecessary to
repair the
equipment)
1.Try to restart
the equipment. If
it is not
recovered,
replace the
control board.
2.Find the
short-circuited
component by
replacing the
component and
repair or replace
it.
3.Use the
service kit to
standardize the
serial cable.
/
SA104
EDC Too Low
SE105 / VCC Too High
/ SA106 VCC Too Low
/ SA107 VCCB Too High
/ SA108 VCCB Too Low
SE109
/
VDD Too High
/
SA110
VDD Too Low
SE111
/
VBB Too High
/
SA112
VBB Too Low
/
SA113
VGC Too High
/
SA114
VGC Too Low
SE115 / VMM Too High
SE116 / VMM Too Low
SE119 / The battery is faulty.
The control board
is faulty.
Try to restart the
equipment. If it is
not recovered,
replace the
control board.
SE120 /
VMM
Connection/Disconnec
tion Self-Test Failure
The control board
is faulty.
Try to restart the
equipment. If it is
not recovered,