EasyManua.ls Logo

Mindray BeneFusion SP3 - Page 40

Mindray BeneFusion SP3
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-4
replace the
control board.
SE121 /
AD Reference Voltage
Failure
The control board
is faulty.
Try to restart the
equipment. If it is
not recovered,
replace the
control board.
/ SA122 Charging Exception
1.Poor battery
connection.
2.The motor is
faulty.
3.The control
board is faulty.
1.Reconnect the
battery.
2.Replace the
battery with a
new one.
3.Replace the
control board.
SE201 /
Main and Auxiliary
Communication
Failure
1.The keypad
board is faulty.
2.The control
board is faulty.
1.Replace the
keypad board
2.Replace the
control board.
SE231 /
EEPROM Failure of
Control Board
The keypad
board is faulty.
Replace the
keypad board.
SE232 /
FLASH1 Failure of
Control Board
The keypad
board is faulty.
Replace the
keypad board.
SE234 /
Monitoring
CPU-EEPROM Read
and Write Errors
The control board
is faulty.
Replace the
control board.
SE251 /
Main control CPU
Self-Test Failure
The control board
is faulty.
Replace the
keypad board
SE252 /
Monitoring CPU
Self-Test Failure
The control board
is faulty.
Replace the
control board.
SE253 /
Main MCU Reset
Failure
The watchdog is
reset, there may
be software
defects.
Feedback to
R&D for
analysis.
SE254 /
Sub MCU Reset
Failure
The watchdog is
reset, there may
be software
defects.
Feedback to
R&D for
analysis.
SE257 / Resource Failure
Software version
or resource file is
burnt incorrectly.
Re-burn the
software version
or resource file.
SE258 / Brand Library Failure
Software version
or brand library is
burnt incorrectly.
Re-burn the
software version
or brand library.
/ SA260 Brand Library Software version Re-burn the