19
BeneVision N22/N19/N17/N15/N12
Quick Reference Guide
a.  Select Alarm Setup   quick key, ECG waveform/numeric tile or Main 
Menu 
 quick key.
b. Select Arrhythmia tab.
c.  Select Threshold tab.
d.   Select arrhythmia threshold and adjust value using the drop down value list.
e. Select the Main Menu 
 quick key or   to return to the main display.
Disable Non-Lethal Arrhythmia Alarms
To disable all arrhythmia alarms except lethal arrhythmias on the N Series bedside 
monitor:
a.  Select  Alarm Setup 
 quick key, ECG waveform/numeric tile or Main 
Menu 
 quick key.
b.  Select Arrhythmia tab.
c.  Select Lethals Only.
d. Select the Main Menu 
 quick key or   to return to the main display.
Accessing Help When Technical Alarms Occur
In the technical alarm list, alarm messages followed by Detail includes help 
information. 
a.  Select the technical alarm information area to enter the Alarms dialog.
b. Select the Technical Alarms tab.
c.  Select the desired alarm. 
d. Select the Main Menu 
 quick key or   to return to the main display.
RESET ALARMS  
To reset an alarm for the duration of the alarm on the N Series bedside monitor, 
select the Alarm Reset 
 quick key.
PAUSE ALARMS/PAUSE ALARM TONES  
To pause alarms or alarm tones on the N Series bedside monitor, select Alarm 
Pause
 quick key to pause all the alarms, or select Audio Pause   quick key 
to pause audible alarm tones.
WARNING: Selecting alarm pause or audio pause may affect current and future