Warnings and Guidelines Calibration and Performance Verification
2 - 2 0070-10-0604-01 Duo™ Service Manual
2.2 Warnings and Guidelines
In the event that the instrument cover is removed, observe the following warnings and
guidelines:
1. Do not short component leads together.
2. Perform all steps in the exact order they are given.
3. Use extreme care when reaching inside the opened instrument. Do not contact exposed
metal parts that may become ‘‘live’’.
4. Read through each step in the procedure so it is understood prior to performing the step.
2.3 Test Equipment and Special Tools Required
• 0-300 mmHg Digital or Mercury manometer with bulb and valve
• 500 cc Test Chamber/Dummy Cuff. P/N 0138-00-0001-03
•DVM
•SpO
2
simulator
• NIBP simulator
• Safety Analyzer (Dempsey model 431 or equivalent)
• Oscilloscope
• Laptop or PC (software upgrade)