Duo™ Service Manual 0070-10-0604-01 2 - 7
Calibration and Performance Verification NIBP Calibration
NOTE: The actual measured values displayed on the Duo monitor
may not compare with the selected target pressure on the
simulator. This test is intended to confirm the REPEATABILITY,
not accuracy, of dynamic NIBP readings. Accuracy can only
be confirmed by performing the NIBP Calibration outlined in
section 2.6 of this manual.