Troubleshooting Guide Repair Information
4 - 4 0070-10-0604-01 Duo™ Service Manual
Monitor Failures
MESSAGE/
PROBLEM REASON SOLUTION
No display after
power-on, power
indicator does not
light.
Bad line fuse Replace fuse
Bad power supply Replace power supply
Bad CPU/Display
board
Replace CPU/Display board
NIBP or SpO
2
will
not function.
CPU/Display board or
module failure.
Isolate and replace defective board/module