Structure and Assembly/Disassembly 9-37
9.3.15.1 Double Sided Mechanism Lock Assembly
1. Refer to Chapter 9.3.15 for details.
2. Remove the screw pad. Unscrew 1 M3 x 10 stainless round inner head screw from the support
arm lock with M3 inner hexagon spanner to remove the support arm lock.
3. Unscrew 4 M5 X 16 stainless steel round inner head screws from the double sided mechanism
lock assembly with M5 inner hexagon spanner to remove the double sided mechanism lock
assembly.
Double sided mechanism lock assembly
Control panel moving mechanism assembly
(FRU)