Image Optimization  5-95 
Operation  Rotate the knob under [E Quality] on the touch screen to select the different THI 
frequency value. 
The adjusting range of basic frequency values can be divided into penetration 
preferred (Pen), general mode (Gen), resolution preferred (Res). Please select the 
frequency according to the detection depth and current tissue features. 
Impacts  Used to optimize the resolution and the penetrability of the image.   
 
Elas Metric 
  Used to adjust the elastography metric. 
  Rotate the knob under [Elas. Metric] to adjust the value on the touch screen. 
The metric includes Young’s modulus E (unit: kPa), shear modulus G (unit: kPa), and 
shear wave velocity Cs (unit: m/s). 
The current elastic modulus or the shear wave velocity (including the unit) appears on 
the top of the color bar. 
 
Scale 
Description  Used to change the maximum scale to make the map related to the color at the top of 
the bar. Optimize the elasto modulus, or mirror the elasto wave velocity to the map. 
Operation 
Rotate the knob under [Scale] on the touch screen. The value on the top of the Map 
changes as the Scale changes. 
The adjusting range: 0 to 29 in increment of 1. 
Impacts  Parts which exceed maximum elasto modulus or shear wave velocity will be mapped 
onto the color on top of the color bar at top-left part of the image. Thus if the color in 
the ROI is mainly the color on top of the color bar, you need to increase the metric 
range. 
 
Opacity 
Description  Used to adjust the opacity feature of the Elasto image. 
Operation  Tap [Opacity] on the touch screen. The adjusting range: 0 to 5 in increment of 1. 
 
Map 
Description 
Used to adjust the color map to achieve the switch between the gray map and the 
color map. 
Operation  Tap [Map] on the touch screen to select the map. E1 is a gray map; E2 and E3 are 
color maps. 
 
ROI Adjustment 
Description 
This feature is used to adjust the ROI position and scale of the lesion detected in STQ 
imaging.