DICOM/HL7 11-5
Name Description
Color Mode
Select the color mode.
If you choose the mix or the grey, RLE/JPEG is unavailable.
The image uses 24 bit when sending the image from the
ultrasound device to the server; it depends on the image when
choosing the mix. The image use 8 bit if the image is captured
in color mode or the image has the tint. All images use 8 bit
when choosing the grey.
Allow Multiframe If SCP supports this function, then select it.
Max Frame Rate
Set the frame range of transferring cin file into DCM multi-
frame file. It is editable to the user.
3D/4D
Set the 3D/4D image transfer mode.
Set the transfer mode for the 3D/4D cine sending.
Normal: use the way that 2D image adopts to send;
Volume: use Enhanced US Volume Storage IOD to send;
Data source: used to obtain 3D/4D image for 4D Viewer.
SR Storage Option To enable or disable structured reporting sending.
Storage mode
Set the storage mode for image and cine file:
Parallel file: save the current file, and is ready for the storage
of the next file.
Parallel frame: send the current frame, and is ready for
sending the next frame.
Add Add the DICOM service to the service list.
Cancel Click to cancel parameter setting.
Update
Select an item in the service list, change the parameters in the
above area, and click [Update] to update the item in the
service list.
Advance
Setting
Encapsulated PDF
Select if to encapsulate PDF format report in DICOM
standard.
It becomes available if SCP supports the function.
Doppler audio Set to save the audio of PW mode.
Transducer Tracking
Files of images that are saved in DCM format through DICOM
or DICOMDIR contain transducer serial number information.
Measurement
If Allis selected, the structured report will contain all the
latest measurement values displayed on the measurement
report.
If Single” is selected, the structured report will contain only
the final measurement value.
Manufacturer
Select the desired manufacturer, and the structured report will
be in accordance with the regulations of the selected
manufacturer.
Strategy Name
Select the desired DICOM strategy.
Service List Delete Click to delete the selected service in the service list.