TransducerControlsandInspection
24 Operator's Manual
3.5 Cleaning and Disinfecting the Transducer
Forcleaninganddisinfectingthetransducer,see"5 CleaningandDisinfection".
WARNING
If you don’t clean and disinfect the transducer, it may become a source of infection.
3.6 Checking after Turning ON the System
CAUTION
If you keep a hot acoustic lens on the body surface, the patient may be burned.
Any of the problems mentioned above indicates that the ultrasonic diagnostic
system or the transducer may be defective.
AfterturningONthepoweroftheultrasonicdiagnosticsystem,performthefollowing
checks.
1. Theacousticlensofthetransducermustnotgenerateabnormalheatwhileitisbeing
used.Thetransducertemperatureshouldbecheckedbyhand.
2. Scanplanecalibration:Ascanplanepositioningcalibrationtestisautomatically
performedwhenthetransducerisconnectedandtheultrasoundsystemisturnedon.
Thiscalibrationcyclelasts5to10seconds.Afterthecalibrationtestiscompleted,
thetransducertemperaturesensorisactivated,andthetransducer temperatureis
displayed,indicatingthetransducerisreadyforuse.
Ifthecalibrationtestofthetransducerfails,(noresponsefromthescanplanebuttons
aftercalibration),re-connectthetransducertorepeatthecalibrationtest.
3. Theimagemustnotbeabnormalwhileturningonthesystem.