CleaningandDisinfection
 Operator's Manual 31
5 
Cleaning and Disinfection
ThissectiondescribesthemethodsandprecautionsforcleaninganddisinfectionofTEE
probes.Aftercompletingeachexamination,clean,disinfecttheTEEprobesasrequired.If
necessary,repeatthecleaning,disinfectionprocessbeforenextuse.
 WARNING
 
●
Keep the control handle and system connector out of any cleaning or disinfection 
solutions. The control handle and cable may be cleaned with a damp cloth, but 
only the distal end of the transducer up to the 100 cm marker on the shaft (70 cm 
for P8-3Ts) may be placed into a disinfection solution.
 
●
Do not use other disinfection methods like Iodine, Steam, Heat or Ethylene Oxide.
 
●
To avoid injury to the patient, you must follow the manufacturer’s 
recommendation for rinsing.
 CAUTION
 
●
To avoid damaging the transducer, the transducer should not be exposed to the 
disinfectant longer than specified to achieve the desired effect, but never longer 
than one hour.
 
●
To avoid damaging the transducer, do not steam autoclave or subject the 
transducer to Ethylene Oxide (ETO).
 
●
To avoid damaging the transducer, do not immerse the transducer in a solution 
containing ethanol.
 
●
After disinfection, rinse the transducer thoroughly with clean water to remove all 
chemical residues. Chemical residues on the transducer may be harmful to the 
human body.
 
●
The efficacy of disinfectants solutions is not guaranteed by MINDRAY. Contact 
the manufacturers for information on the activity of the products.
NOTE:
 
●
Aftertheexamination,wipeofftheultrasoundgelthoroughly,otherwise,theultrasound
gelmaysolidifyanddegradetheimagequalityofthetransducer.
 
●
Donotpermitthetransducertobecomeoverheated(morethan55°C)duringcleaning
anddisinfections.Hightemperaturemaycausethetransducertobecomedeformedor
damaged.
 
●
Cleantheprobethoroughlyinaccordancewiththecleaningprocedurebefore
disinfection.