CleaningandDisinfection
32 Operator's Manual
5.1 Before Processing
Thisstepistoremovetheultrasoundgelorothervisibledirt.
1. Wearapairofglovestopreventinfectionthroughthewholecleaning,disinfection
process.
2. Disconnecttheprobefromthesystem.Ifthesheathisused,takeoffthesheathand
disposeitasdirectedbythehospital.
3. Wipeawaytheultrasoundgelorothervisibledirtonthesurfaceoftheprobebyusing
adamppieceofdisposablelint-freesoftclothortissue.
5.2 Cleaning
OnlythefollowingcleanersarerecommendedbyMindraytocleantheTEEtransducers.
Forthebiologicaleffectivenessandthecorrectuseofthecleaners,seetheinformationof
thecleaner’smanufacturer
.
5.2.1 Compatible Detergents
Cleaner Active ingredient Concentration Applicable for
Cidezyme/Enzol Proteolyticenzymes < 5% P7-3T
P7-3Ts
P7-3TU
P7-3TE
P8-3Ts
P8-2Ts
P8-2TU
EMpower Proteolyticenzymes < 2%
Metrizyme Proteolyticenzymes < 2%
Neodisher
MediCleanforte
Trisodiumnitrilotriacetate 5-10%
2,2-Iminodiethanol
diethanolamine
1-2%
Prolystica2x
conc.Enzymatic
Presoak&cleaner
Ethanolamine 1-5% P7-3T
P7-3Ts
P7-3TU
P7-3TE
P8-3Ts
P8-2Ts
P8-2TU
Protease 0
.
1-1%
Ethoxylatedalcohol 1-5%
Polyalkyleneglycol 1-5%
Glycerine 1-5%
WIP’ANIOS
CLEAN’UP
Associationofsurfactants,
excipients
/ P8-2Ts
P8-2TU
5.2.2 Cleaning Procedure for TEE Probe
1. Selectanappropriatecleanerincludecleaningwipesandsprays,milddetergents,
enzymaticcleanersandspeciallydesignedenzymaticsponges.
2. Followthemanufacturer’sinstructionstoprepareandusethecleaner.Selectan
appropriatemethod:
Soaking:Soaktheprobeinthecleanersolutionforatleast1minuteorfollowthe
manufacturer'sinstructions.Wipeandwashtheprobesurfacegentlybyusinga
pieceoflint-freesoftclothorsoftspongeuntilnodirtisvisible.Whennecessary,
washthelocatinggrooveandotheritemsbyusingdisposablecottonswabs.
Avoidusingabrushtowashthelensbecauseitmaydamagetheprobe.