EasyManua.ls Logo

Mindray P8-2Ts - Emergency Retraction; After Examinations

Mindray P8-2Ts
51 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Examination
Operator's Manual 29
CAUTION
To avoid damaging the transducer, do not deflect the distal tip of the transducer by
direct application of force. Use the deflection wheels for this task.
4.3.3 Emergency Retraction
Ifthetransducertipshouldgetjammedinadeectedpositioninsidethepatient,andall
attemptstoreleasethedeectedtipshouldfail,followthefollowingproceduretoassurea
saferetractionofthetransducer.
1. Disconnectthetransducerfromtheultrasoundsystem.
2. Atanaccessiblelocationbetweenthetransducerhandleandthepatient,cuttheentire
endoscopeshaft,includingallinternalwiring,usingheavydutycuttingpliersoranother
suitable tool.
Thedeectionmechanismisnowreleasedandthetransducermaybesafelyretracted.
4.4 After Examinations
Aftertheexaminationiscompleted,turnOFFtheultrasonicdiagnosticsystem,removethe
bite-guardanddisconnectthetransducer,andthencleananddisinfectthetransducer.
Afterdisinfectingthetransducer,conrmthatthetransducerisingoodconditionandstoreit
inasuitableplace.
Fordetailsaboutcleananddisinfect,see"5 CleaningandDisinfection".

Table of Contents

Related product manuals