Specications
38 Operator's Manual
P8-3Ts
Centerfrequency 5.0MHz
Transducertype Phasedarray
Numberofelements 48
Aperture 7.25 mm
Focus 30 mm
Scanplanerotation 0
180°
Temperatureprecision Temperaturerange:17.5~50°C,precision:±1°C,ofwhich,
whenthetemperatureis43°C,theprecisionis±0.5°C
Tipdeection Up/Down:120°/90°
(Deviation:-10%mm,theupperlimitisnotcalculated)
Leakagecurrent/Dielectric
strength
MeettherequirementsofIEC60601.
Biocompatibility AllexternalmaterialsofthemultiplaneTEETransducer
havepassedrelevantbiocompatibilitytestsaccordingISO
10993.
P8-2Ts/P8-2TU
Shaftlength 108cm(deviation:±10%)
Mechanical
dimensions
Shaftexternal
diameter
10.5mm(deviation:+10%,thelowerlimitisnotcalculated)
Diameter of
theinexible
distalpartofthe
transducertip
17.5mm(deviation:+5%,thelowerlimitisnotcalculated)
Maximum
insertion portion
diameter:
≤17.5mm
Centerfrequency 5.0MHz
Transducertype Phasedarray
Numberofelements 64
Aperture 9 mm
Focus 60 mm
Scanplanerotation 0
180°
Temperatureprecision Temperaturerange:17.5~50°C,precision:±1°C,ofwhich,
whenthetemperatureis43°C,theprecisionis±0.5°C
Tipdeection Up/Down/Left/Right:≥120°/45°/45°/45°
(Deviation:-10%mm,theupperlimitisnotcalculated)
Leakagecurrent/Dielectric
strength
MeettherequirementsofIEC60601.
Biocompatibility AllexternalmaterialsofthemultiplaneTEETransducer
havepassedrelevantbiocompatibilitytestsaccordingISO
10993.

Table of Contents