2 - 10 Accutorr 7 Operator’s Manual
indicates alarms are reset.
indicates alarm sounds are turned off.
3. Parameter and waveform area: displays parameters and waveforms.
4. Manual input area: manually input physiological related values. This area does not
display by default. Refer to 4.2 Manually Input Patient Data for additional
information.
5. Menu QuickKeys
â—† Main: Accesses the [Main] screen to configure the monitor, or quickly
returns to the main screen.
◆ Scoring: Accesses the [Scoring] screen to evaluate a patient’s condition.
Refer to 12 Clinical Scoring.
â—† Patient List: Accesses the [Local Patient List] and [ADT Database] screen to
admit a patient stored in the monitor or ADT database. Refer to
4.1.3 Admitting a Patient from [Patient List].
â—† Review: Displays the spot check trends, continuous trends, Scoring Review,
and graphic trends. Refer to 4.5 Reviewing Patient Data.
â—† Save: Accesses the [Results] screen to manually save patient data. Refer to
4.4 Manually Save Patient Data.
6. Network and USB connection area
Display CMS information or Spot Check mode. For details for how to configure
CMS, refer to 14.3 Connecting the monitor to the CMS.
Indicate the connection of network or USB to this monitor.
7. Battery status: indicates the status of the battery. For details, refer to 15 Battery.
8. System time
9. Tabular trend area. This area displays only in Trend screen mode.
10. NIBP list area. This area displays only in NIBP List screen mode.
indicates monitor is successfully connected to a wired network.
indicates the wireless function (optional) is working.
indicates monitor has failed to connect a wired network.
indicates the wireless function (optional) is not working.
indicates a USB drive is inserted.