12-24 Probes and Biopsy
Glutaraldehyde-based sterilant:
Chemical
name
Trade name
Procedures
Glutaraldehyde
(2.2-2.7%)
Cidex
Activated
Glutaraldehyde
Solution
Please refer to the instructions provided by the
manufacturer of the solution for details.
Soak the transducer into the activated solution for 10
hours (20-25)
Hydrogen Peroxide and Peroxyacetic Acid -based sterilant:
Trade Name
Chemical Name
Procedures
Minncare
®
Cold
Sterilant
22% Hydrogen
Peroxide
4.5% Peroxyacetic Acid
Dilute the sterilant with dsterilized purified
water (1:20).
Immersed time: 11 hours.
Temperature: 20-25.
Please refer to the instructions provided by
the manufacturer of the solution for details.
Refer to the instructions provided by the chemical manufacturer concerning
concentration of the solution, and method of disinfections and dilution. Note that the
glutaraldehyde disinfectant solution needs an activating solution.
Rinse the needle-guided bracket thoroughly with sterile water to remove all chemical
residues on it.
Wipe off the water on the needle-guided bracket with sterile cloth or gauze after
rinsing it.
STERRAD 100S low-temperature hydrogen peroxide gas plasma sterilization system
Chemical name
Trade name
Procedures
Hydrogen
peroxide gas
plasma
Hydrogen
peroxide vapor
Please refer to the instructions provided by the
producer of the solution for details.
Refer to the instruction of STERRAD 100S sterilizing system provided by the
manufacturer for operation instructions and cautions.
The STERRAD 100S low-temperature hydrogen peroxide gas plasma sterilization
system is available for metal needle-guided brackets.
High-pressure steam sterilization (only applicable for metal guided-bracket)
Autoclaving (moist heat) 121C for 20 minutes.
NOTE:
The high-pressure steam/ immersion sterilization do not affect the bracket duration
life, and the duration life is affected by the daily application of the bracket. Please
check the appearance of the bracket before using.

Table of Contents

Other manuals for Mindray Z5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Z5 and is the answer not in the manual?

Mindray Z5 Specifications

General IconGeneral
BrandMindray
ModelZ5
CategoryMedical Equipment
LanguageEnglish

Related product manuals