A-2 Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
1- Power Cord Plug
TEST PROCEDURE
The Power Plug
The Power Plug Pins
No broken or bent pin. No discolored pins.
The Plug Body
No physical damage to the plug body.
The Strain Relief
No physical damage to the strain relief. No plug
warmth for device in use.
The Power Plug
No loose connections.
The Power Cord
The Power Cord
No physical damage to the cord. No deterioration
to the cord.
--For devices with detachable power cords, inspect
the connection at the device.
--For devices with non-detachable power cords,
inspect the strain relief at the device.