A-6  Electrical  Safety  Inspection 
ELECTRICAL SAFETY INSPECTION 
4- Protective Earth Resistance   
metal area.   
3)  Press shortcut key 3. The Protective Earth Resistance test is displayed.   
4)  Press SOFT  KEY 3 to select a test  current (1AMP, 10AMP, or 25AMP). The selected test 
current is displayed in the upper right corner of the display.   
 
5)  Press START TEST to start the test. The test current is applied while resistance and current 
readings are taken. This takes approximately 5 seconds. 
6)  Press the print data key at any time to generate a printout of the latest measurement(s). 
 
When "Over" is displayed for Ohms, this signifies that a valid measurement was not obtained 
because either an open connection was detected or that the measurement was not within range. 
Readings greater than 9.999 Ohms will be displayed as Over.   
Once it reaches the limitation, stop using equipment. Check the protective earth connection 
between Protective Earth terminal and Protective Earth Metal enclosure and equipotential terminal; 
Retest and inform the Customer Service Engineer for analysis and disposal if still fail. 
ALL COUNTRIES    R = 0.2Ω Maximum