Electrical  Safety  Inspection  A-15 
 
ELECTRICAL SAFETY INSPECTION FORM 
(Class I equipment) 
Overall assessment: 
Opened repair type, not modify the power part 
including transformer or patient circuit board 
Test item: 1, 2, 3, 4, 5, 9 
Opened repair type, modify the power part 
including transformer or patient circuit board 
Test item: 1, 2, 3, 4, 5, 6, 7, 8, 9 
Measurement equipment /SN: 
Device Enclosure and Accessories 
Protective Earth Resistance 
Max:     
NC: 300μA(refer to 
UL60601-1) * 
NC: 500μA(refer to   
IEC60601-1) * 
SFC: 1000μA 
Single Fault condition(SFC) 
Max: 
BF applied part: 
NC:100μA, SFC: 500μA 
Single Fault condition(SFC) 
Mains on Applied Part Leakage 
Max: 
BF applied part: 5000μA   
Patient Auxliary 
Leakage Current 
Max: 
BF applied part: 
NC:100μA, SFC: 500μA 
Single Fault 
condition(SFC) 
Note: 
 The equipment which sell to America shall comply with the requirement of UL60601-1, 
others shall comply with the requirement of IEC60601-1. 
Name/ Signature: ____________________ Date: ____________________________