Operation Manual Maintenance
10-1
10. MAINTENANCE
It is recommended to periodically inspect the unit, both visually and physically, to insure
optimum and safe performance. This inspection should be scheduled as a routine maintenance
procedure, done with the Sonicator
power switch in the OFF position, and with the unit
unplugged from the AC power source.
Long exposure to acids or caustics results in corrosion of metal parts or components. Check
the generator, convertor, and cables periodically for any signs of rust or discoloration. If
discoloration is found, move the Sonicator
away from the source of the contaminant.
Examine the condition of the high voltage cable that attaches the convertor to the generator.
Inspect the wire insulation for damage, such as wear, burning from hot plate contact or
breakage from extended use or rough handling.
Inspect the cable connectors by gently pulling on the wire while holding the metal portion of the
connector. The cable connectors and rubber boot protectors at both ends of the cable should
be tight to the wire, with no movement possible and no cracks or frayed ends visible. Do not
subject the cable ends to severe bending loads while performing these tests. Order a
replacement cable immediately if your cable does not pass the above inspection. Should the
convertor/cable assembly be subjected to misuse, such as dropping or a severe pulling force
on the wire itself, the cable must be inspected as above.
DO NOT USE A CABLE WITH BROKEN END CONNECTIONS, EXPOSED WIRES OR
FRAYED INSULATION. HIGH VOLTAGE IS PRESENT IN THE CABLE AND MAY POSE A
SHOCK HAZARD. DO NOT TOUCH THE CONVERTOR ASSEMBLY UNTIL THE POWER
SWITCH OF THE GENERATOR IS OFF AND THE UNIT IS UNPLUGGED.
In general use, the cable assembly should not be used to carry the convertor or pull it toward
the user. Make certain the cable always has slack and is never tensioned. If necessary, move
the generator or convertor assembly closer to one another to accomplish this. If this is not
possible, contact Misonix to obtain a longer cable.
Should the Sonicator
stop functioning, shut the unit off and inspect the cable as above
before any other action is taken.
If the operator is in doubt as to the condition of the unit, call 800-645-9846 to speak with a
customer service representative for prompt attention.
Cleaning Instructions
The generator and convertor may be cleaned using an acid-free cleaning solution (i.e. glass cleaner).
Horns and probes should be cleaned using isopropyl alcohol.