85
No. 99MAH056A
8 Error Displays and Countermeasures
No. Error name LCD display Cause Countermeasures
40
Sensor
Contamination
Detection Error
Err 40
A sudden change
in temperature
may create
condensation
on the detector,
or it may be
contaminated by
other sources.
Although this may occur
during high-speed
plunger movement,
thereisnoe󰀨ecton
measurement.
Use the product as is.
y If it occurs while the
plunger is not moving,
turn the power OFF
and allow the product
to adapt to the
temperature for about
2 hours.
y If it does not recover
after adapting to the
temperature, repair
is required. Please
contact the agent
where you purchased
the product or a
Mitutoyo sales/service
representative.
(
"12O󰀨-Site
Repairs (Subject
to Charge)" on
page 105)
41
Internal
Connection
Error
Err 41
There is a problem
with the internal
connection.
The product may be
faulty. Please contact
the agent where you
purchased the product or
a Mitutoyo sales/service
representative.
(
"12O󰀨-SiteRepairs
(Subject to Charge)" on
page 105)
50
Serial
Communication
Forced Error
Err 50
A serial
communication
command [B7]
(forced error
display) is
received.
Send a serial
communication
command [B8] (error
reset).

Table of Contents