EasyManua.ls Logo

Mitutoyo MICROPAK-2LA - Chapter 5 Error Messages; 5.1 Error Messages

Mitutoyo MICROPAK-2LA
36 pages
Print Icon
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error
Messages
5.1
ERROR
MESSAGES
The MICROPAK-2LA generates four. different error messages as listed
in
the
following table.
Table
5-1.
Error Messages
The error messages
El
and
E2
may be cleared by pressing the [CLEAR] key.
Error Indication
E
1
E2
E3
E4
Only the [RESET] key can clear error message E3.
Description
EIA-232D communication error.
Geometric calculation error.
Overspeed or overflow of the counting function.
Abnormality (noise or incorrect voltage levels)
in
the input
signals.
MICROPAK-2LA system error.
When the
unit
generates error message
E4,
please contact your nearest Mitutoyo
service office as listed on the back cover of this manual for assistance.