2.3.2 Selecting Chosen Wavelength
1. From the power measurement mode, press the
menu button until "more" appears. Press "more".
2. Press "select" until "wavelength" is highlighted.
Press "go".
3. Press "change" then "up" and "down" to select the
first wavelength. Repeat steps 2 and 3 for other
wavelengths desired. Up to 6 wavelengths may be
selected.
2.3.3 Setting Startup Configuration
1. From the power measurement mode, press the
menu button until "more" appears. Press "more".
2. Press "select" until "configure" is highlighted.
Press "go".
3. Now press "value" to select filter in or out to be the
default.
4. Press "select" and "value" to choose the manual
power range you wish to be the default or choose
autorange.
5. Now press "select" and "value" again to choose the
startup laser wavelength. For further details on
configuration See Section 5.3.
2.3.4 Power Measurement
Warning:
Do not exceed maximum head limits for power,
energy, power density and energy density as
listed in tables 6 and 7 section 9.2 Head
Specifications. Otherwise, there is a risk of
damaging the absorber.
To expand the bargraph scale 10% about the
present reading:
1. From the bargraph power measurement screen
press the menu button. Then press "zoom". Now
press the menu button twice again.