13
2.4.2 To set type of laser being used
1. From bargraph measurement screen, press the
menu button twice and press "laser" until the
correct laser type or wavelength is displayed.
2. Return to bargraph screen by pressing the menu
button again.
2.4.3 To set Laser Pulse and Average
1. From the bargraph measurement screen press the
menu button twice then press "setup".
2. Press "value" to choose the short pulse or long
pulse setting depending whether the pulses of
your laser are less than or greater than the short
setting.
3. Press "select" then "value" to choose the time
period over which you wish to average energy
readings. Press "exit".
4. Save configuration as described in section 2.4.4.
Warning:
Incorrect readings will result if pulse length is
not set up correctly.
Note:
Some heads have only one time setting for all
pulse lengths. In that case “N.A.” appears.
2.4.4 Setting Startup Configuration
1. From the bargraph energy measurement mode,
press the menu button until "more" appears. Press
"more".
2. Press "select" until "config" is highlighted. Press
"go".
3. Press "value" to choose "power" or "energy" for the
desired startup screen.
4. Press "select" until the laser wavelength is
highlighted. Now press "value" to select the laser
type you wish to be the default.