EasyManuals Logo

MKS Ophir Nova User Manual

MKS Ophir Nova
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
8
To choose power or energy measurement:
1. To go from the bargraph power measurement
screen to the energy measurement screen, press
the menu button then press "energy".
2. Alternatively, you can press the leftmost soft key
directly from the bargraph screen to go to energy
measurement.
3. To go from the energy measurement screen to
power measurement press the "power" soft key.
2.2.2 Setting and Saving Startup Configuration
1. From the power measurement mode, press the
menu button located on the right side until "more"
appears. Press "more".
2. Press "select" until "configure" is highlighted.
Press "go".
3. Press "value" to choose "power" or "energy" for the
desired startup screen.
4. Press "select" until the laser type is high-lighted.
Now press "value" to select the laser type you
wish to be the default.
5. Press "select" and "value" to choose the manual
power range you wish to be the default or choose
autorange.
6. Now press "select" and "value" again to choose the
default energy range.
7. Press "exit" then "all" to save all present settings.
Further details on configuration in section 4.3.
2.2.3 Power or Single Shot Energy Measurement
Warning:
Do not exceed maximum head limits for power,
energy, power density and energy density as
listed in tables 6 and 7 in section 9.2. Otherwise,
there is a risk of damaging the absorber.

Table of Contents

Other manuals for MKS Ophir Nova

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MKS Ophir Nova and is the answer not in the manual?

MKS Ophir Nova Specifications

General IconGeneral
BrandMKS
ModelOphir Nova
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals