+1(323)904-3000 38 www.mobiusa.com
Guidance and manufacturer's declaration – Electromagnetic
immunity – for all equipment and systems
Guidance and manufacturer´s declaration – Electromagnetic
immunity
The Smart Bluetooth Ear + Forehead DualScan™
Thermometer is intended for use in the electromagnetic
environment specified below. The customer or the user of
The Smart Bluetooth Ear + Forehead DualScan™
Thermometer should ensure that it is used in such an
environment.
Immunity
test
IEC
60601
test
level
Compliance
level
Electromagnetic
environment guide
Electrostatic
discharge
(ESD)
IEC 61000-
4-2
6kV
contact
±8 kV
air
±6 kV
contact
±8 kV air
Floors should be
wood, concrete, or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least
30%.
Power
frequency
(50/60 Hz)
magnetic
field
IEC 61000-
4-8
3 A/m 3 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.