Guidance and manufacturer's declaration – Electromagnetic
immunity – for all equipment and systems that are not life-
supporting
Guidance and manufacturer´s declaration – Electromagnetic
immunity
The Smart Bluetooth Ear + Forehead DualScan™
Thermometer is intended for use in the electromagnetic
environment specified below. The customer or the user of
The Smart Bluetooth Ear + Forehead DualScan™
Thermometer should ensure that it is used in such an
environment.
Immunity
test
IEC
60601
test
level
Compliance
level
Electromagnetic
environment guide
Radiated
RF IEC
61000-4-
3
3 V/m
80
MHz to
2.5GHz
3 V/m
Portable and mobile RF
communications
equipment should be used
no closer to any part of the
DualScan™ Thermometer,
including cables, than the
recommended separation
distance calculated from
the equation applicable to
the transmitter's
frequency.
Recommended separation