780
Pulse
modulation
217Hz
810
800-960
GSM 800/900,
TETRA 800,
IDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
18Hz
2 0.3 28
870
930
1720
1700-
1990
GSM 1800,
CDMA 1900,
GSM 1900,
DECT, LTE
Band 1, 3, 4,
25; UMTS
Pulse
modulation
217Hz
2 0.3 28
1845
1970
2450
2400-
2570
Bluetooth,
WLAN, 802.11
b/g/n, RFID
2450, LTE
Band 7
Pulse
modulation
217Hz
2 0.3 28
5240
5100-
5800
WLAN 802.11
a/n
Pulse
modulation
217Hz
0.2 0.3 9 5500
5785
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the
transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1m. The
1m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because
while it does not represent actual modulation, it would be worst case.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic
Environment – Guidance
Radiated RF 3 V/m
Complies Not Applicable
IEC 61000-4-3
80 MHz-2.7 GHz
80% AM at 1 kHz
NOTE: The EMISSIONS characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which
CISPR 11 class B is normally required) this equipment might not offer adequate Protection
to radio-frequency communication services. The user might need to take mitigation
measures, such as relocating or re-orienting the equipment.