Trueprep AUTO v2
8.3 Maintenance
8.3.1 Disposal of consumables
Discard the used cartridges, trays, reagent bottles and other consumables in 0.5% Replaceable
Sodium Hypochlorite. Let the items remain submerged in the solution for 30 minutes and then
dispose as per the medical waste guidelines in your region.
8.3.2 Cleaning
The surface of the device should be cleaned using freshly prepared 0.5% Sodium Hypochlorite
solution and a clean cloth or tissue. This should be done once everyday and/or when the device is
moved to a different work space.
8.3.3 Spills
If you notice a spill in the tray placed in the cartridge holder, dispose the tray. Clean Replaceable
the insides of the cartridge holder with clean cloth or tissue or swab dipped in 70% Isopropyl
Alcohol (IPA). Wait for 5 minutes before inserting a new tray into the holder.Replaceable
8.3.4 Disposal of Instrumentation
Do not Dispose off Instrument as unsorted municipal waste. Contact Molbio for collection of
instrument.