5014166 D 51
3
Environment Control and Fluidics
Options Maintenance
It is important to regularly clean the environmental enclosure, including
specific fluidics and environmental components. Components can come
in contact with biological, chemical, and toxic agents. Therefore, all
cleaning procedures should be handled with care. Molecular Devices
recommends that you wear powder-free gloves at all times when you
access the internal components of the enclosure. For additional
information about cleaning the system, see the ImageXpress Micro
Widefield High Content Imaging System User Guide or the ImageXpress
Micro Confocal High Content Imaging System User Guide.
The following sections provide information on how to clean fluidics
components. If you have any further questions about specific cleaning
procedures, contact Technical Support. See
Obtaining Support on page 7.
This chapter contains the following sections:
Carbon Dioxide Tubing on page 52
Cleaning the Water Reservoir on page 53
Waste Disposal Box and Plate Sealing Ring on page 54
CAUTION! Never use an autoclave to clean any instrument
components.
ImageXpressOptionsUserGuide.book Page 51 Wednesday, September 9, 2015 3:54 PM