Chapter 3: Objective Maintenance
5063669 E 19
Adjusting an Objective Correction Collar for a Plate
To adjust an objective correction collar for a slide:
1.
In the CellReporterXpress software, on the Home page, click Acquisition.
2. Click Add Protocol.
3. Click New Slide Acquisition.
4. In the Available Acquisition Devices list, select the instrument.
5.
On the left side of the screen under Steps, click Acquisition Settings.
6.
On the right side of the screen under Tools, click Slide Format.
7. In the Slide Format list, select the slide format.
8.
On the right side of the screen under Tools, click Objectives.
9. In the Objectives list, select the objective.
If a correction collar adjustment is required, the software displays the recommended
setting.
Note: When using the environmental control cassette with a 40x objective, you
must adjust the objective correction collar an additional 0.7 mm above the
recommended setting to account for the thickness of the cassette glass bottom.