Fig 22
When you enter the overview of the parameters in magnetic absorption, you can alter several settings.
It is not advised to alter these, as any change can have a significant effect on operation efficiency. Only
trained personnel can alter these settings. The parameters are explained below:
Segments: The setting range is 0 ~ 5, it can stop magnetic absorption per segment, the
magnetization function will be off if set it to 0.
Cycle times: Times the magnetic absorption cycle is repeated.
Mag.speed: The speed of magnetic absorption when the magnetic rod moves under liquid
level. 1 is the slowest while 10 the fastest.
Lip-lvl: The standing time for magnetic rods when they pull away, closing the liquid level after
finishing magnetic absorption. This standing time is set to ensure the magnetic beads stay
bound to the magnetic rods and sleeve, and the beads do not fall off due to liquid surface
tension.
Anti-splash: The standing time for the magnetic rods, pulling away from the liquid level after
finishing magnetic absorption. This standing time is set to prevent cross contamination caused
by liquid splashing or drops which can fall back into the liquid shortly after pulling away.
N (1-5) Segment time: Independent magnetic absorption time of each segment, the maximum
time can reach 999 seconds.
Estimated: The estimated magnetic absorption time of the software. It can only be displayed on
the next entry after exiting the interface.