Panorama Exposures
48
CLASS 2 LASER PRODUCT: A class 2 laser is used for
the positioning beams.
The laser beams could damage the eyes.
Do not stare into the positioning beams. Warn the patient
not to look at the positioning beams.
Warn the patient not to move during emission (while the
melody is sounding). If the patient moves, he could be hit
by the arm or the image might be ruined.
* The “Double-Ready
*1
Function” reduces the length of time that the emission button needs to be held down.
Press the Ready key a second time after completing patient positioning in Ready mode; a two-toned beep
will sound and the arm will move to its exposure start position. The Ready LED will light up again and the
Image layer beam will go out.
* In Double-Ready
*1
mode, X-ray emission starts as soon as the emission button is pressed.
* In Double-Ready
*1
mode, other operation keys and switches are disabled except for the Ready key and Lift
keys.
* In Double-Ready
*1
mode, press the Ready key again to return to the patient positioning Ready mode.
Usage Note
If the patient moves out of position after the unit is set for Double-Ready
*1
, press the Ready key again to
return the unit to its normal Ready setting and re-position the patient.
*1
The Double-Ready function is not enabled by default. To enable the Double-Ready function, contact your local dealer or
J. MORITA OFFICE.