EasyManua.ls Logo

Motic EasyScan One - Page 30

Motic EasyScan One
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
● Z Compensation Calibration
Z Compensation calibration helps to compensate the error of Z axis movement.
Current compensation (um): Display the calibration result of the Z axis compensation under 40x
objective.
Calibrate: To start Z axis compensation calibration.
b. Macro Camera Calibration
Below figures are for video image calibration and slide image position calibration.
Video image:
Before calibration, make sure you have well loaded the tray with slide which has shape tissue and label
on it. Then please follow the calibration wizard. (Fig. B)
● Click “Prepare Tissue Exposure” on the calibration wizard window, then tray will step back. You
will find the marco camera’s field of view is blank.
● Ensure the tray does not exist on the device.
● Adjust the exposure bar to let the histogram close to red bar (Fig. A)
● Click “Set Tissue Image” on the calibration wizard window, then system will automatically
complete exposure, background balance and auto white balance setting.
● Adjust the brightness to make the peak of histogram, which inside the limit range, at right side.
(Fig. C)
● Put the tray on the device.

Related product manuals