− Then the program automatically goes to the internal calibration
procedure which is signalled by the following message:
− After completion of the calibration process program returns
to the weighing mode.
• Calibration triggered off by timer
− The time condition for subsequent automatic calibration is 3
hours. It means that, when no other triggering factor appear,
the calibration will appear every 3 hours;
− The calibration procedure triggered off by the time change starts
with checking which is signalled by the following message:
− Then the program automatically goes to the internal calibration
procedure which is signalled by the following message:
− The program for non-verified scales has a parameter for setting
a maximal time interval between subsequent internal calibration.
Procedure:
− Enter the submenu <P6.CAL> and then: