17.1.2. Automatic internal calibration
The automatic calibration process can be triggered off by 3 different
factors:
Calibration after powering up
After performing the start procedure the scale program starts
to check stability conditions for the calibration process and
displays the following message:
Then the program automatically goes to the internal calibration
procedure which is signalled by the following message:
After completion of the calibration process program returns
to weighing mode.
Calibration triggered off by temperature changes
The scale has been equipped in the temperature monitoring
system;
Temperature triggers off calibration every time when the internal
system measures the temperature change greater than 3°C;
The calibration procedure triggered off by the temperature
change starts with checking which is signalled by the
following message: