Table 1-2. Typical Chillgard LS / R-11 Cross-Sensitivity Response Data
GAS CONCENTRATION (PPM) EQUIVALENT PPM R-11
Acetone 100 14
Methyl Ethyl Ketone 100 32
Methanol 1000 130
iso-Propanol 500 400
Methylene Chloride 100 0
Xylene 100 0
Ammonia 50 0
R-113 100 25
R-11 100 100
R-22 100 30
R-12 983 490
R-134 A 100 20
R-123 100 3
Propane 0.6% 18
Ethylene 500 20
TEST CONDITIONS: Calibrated 0-1000 PPM R-11 in N
2;
Temperature: 25
O
C
Table 1-3. Typical Chillgard LS / R-12 Cross-Sensitivity Response Data
GAS CONCENTRATION (PPM) EQUIVALENT PPM R-12
Acetone 100 3
Methyl Ethyl Ketone 1000 25
Methanol 1000 2
Methylene Chloride 1000 14
Trichloroethylene 1000 167
Ethyl Acetate 1000 22
Xylene (ortho-Xylene) 1000 5
R-113 100 35
R-11 100 5
R-22 100 1
R-12 100 100
R-134 A 100 2
R-114 100 45
TEST CONDITIONS: Calibrated 0-1000 PPM R-12 in N
2
; Temperature: 25
O
C
1-5