Table 12 Digital Sensor Default Settings
Gas
(Code)
Threa
d
Type
3
TruCa
l
Range
Default
2
Display
Resolutio
n
5
Unit
Defaul
t
Warn
Defaul
t
Warn
Defaul
t AO4
(mA)
Alarm
Defaul
t
Alarm
Defaul
t AO
4
(mA)
Alarm
Action
Default
Cal
Level
Defaul
t
Alar
m
Min
Alar
m
Max
Rang
e Min
Rang
e max
Cal
Level
Min
Cal
Leve
l Max
Ammonia
(D40)
Coarse 0-100 1 PPM 10 5.6 20 7.2 Increasing
Non-Latch
25 2 100 0-10 0-100 1 FS
1
Ammonia
(D41)
Coarse 0-1000 10 PPM 100 5.6 200 7.2 Increasing
Non-Latch
300 100 1000 190 1000 290 FS
1
Carbon
Monoxide
(D10,
D14) *
Fine 0-100 1 PPM 15 6.4 30 8.8 Increasing
Non-Latch
60 15 1000 0-10 0-1000 5 FS
1
Carbon
Monoxide
(D11) *
Fine 0-500 1 PPM 50 5.6 150 8.8 Increasing
Non-Latch
300 15 1000 0-10 0-1000 5 FS
1
Carbon
Monoxide
(D12) *
Fine 0-1000 1 PPM 100 5.6 300 8.8 Increasing
Non-Latch
400 15 1000 0-10 0-1000 5 FS
1
Carbon
Monoxide
(D36/D3
9) *
Fine 0-100 1 PPM 15 6.4 30 8.8 Increasing
Non-Latch
60 15 1000 0-10 0-1000 5 FS
1
Carbon
Monoxide
(D37) *
Fine 0-500 1 PPM 50 5.6 150 8.8 Increasing
Non-Latch
300 15 1000 0-10 0-1000 5 FS
1
Carbon
Monoxide
(D38) *
Fine 0-1000 1 PPM 100 5.6 300 8.8 Increasing
Non-Latch
400 15 1000 0-10 0-1000 5 FS
1
37
US 4 Operation